Dashboard/Back office/Assets & Biomed

Assets & biomedical equipment

Asset register, preventive maintenance (PM) and calibration, breakdown tickets with downtime, and depreciation. Biomedical engineer: Zeeshan Raza.

Registered assets
312
68 biomedical · book value Rs 30.97 M
PM compliance (90 d)
91%
3 overdue · target ≥ 95%
Open breakdowns
2
1 critical (lab analyser)
Downtime this month
14.5 h
Critical equipment · MTTR 4.8 h
BT-26-034 · Mindray BS-480 chemistry analyser down since 05:40 (6 h)

Impact: 186 chemistry samples rerouted to the backup Cobas c311 · routine chemistry TAT +2.4 h · 31 B2B clients notified that reports are delayed. Engineer from Premier Agencies on site 11:30.

Lab TAT →
TagAssetLocationSerialVendorWarrantyAMC / CMCCriticalityStatus
CW-LAB-0007Mindray BS-480 chemistry analyserLab · ChemistryBS4-2203117Premier AgenciesExpired Mar 2025CMC till 31 Mar 2027CriticalDown
CW-LAB-0003Sysmex XN-550 haematologyLab · HaematologyXN-11902Sysmex distributorTill Jun 2027—CriticalIn service
CW-RAD-0001DR X-ray system 500 mADHA · ImagingDRX-88412Imaging vendorTill Jan 2027AMC (PNRA-authorised)CriticalIn service
CW-RAD-0004Ultrasound (colour Doppler)DHA · ImagingUS-5521-KImaging vendorExpiredAMC till Dec 2026HighIn service
CW-DC-0002Autoclave 75 L (class B)Day-care · CSSDAC75-09981Local supplierExpired—HighUnder repair
CW-PH-0011Pharmacy vaccine fridgeDHA · PharmacyVF-2290Local supplierTill Aug 2027—HighIn service
CW-GEN-0001Diesel generator 60 kVADHA · UtilityGEN-60-4410Local supplier—Quarterly serviceMediumIn service
October 2026 · week 12–18PMCalibrationOverdueDone
Mon 12Sysmex daily maint.Generator weekly run
Tue 13Autoclave Bowie-Dick validationBS-480 weekly maint.
Wed 14Pipette calibration (12)
Thu 15USG probe check
Fri 16Centrifuge RPM/timerAC filters
Sat 17Fridge thermometer cert.
Sun 18
AssetTaskFrequencyLast doneNext dueByStatus
Autoclave 75 LBowie-Dick + spore testDaily / weekly06 Oct13 OctCSSD techOverdue
DR X-rayQC (kVp, mA, HVL) + leakage surveyAnnual (PNRA)22 Jan 202622 Jan 2027PNRA-authorised service providerValid
Mindray BS-480Photometer & ISE calibrationMonthly14 Sep14 OctLab + vendorDue
Pipettes (12)Gravimetric calibration6-monthly14 Apr14 OctZeeshan RazaScheduled
Defibrillator (day-care)Energy output testQuarterly02 Jul02 OctZeeshan RazaOverdue
PNRA note

Radiation equipment (X-ray, OPG, CT) may only be serviced by PNRA-authorised engineers. Service reports, QC results and staff TLD dose records are kept here for licence renewal.

Reported→Assigned→Under repair→Resolved
TicketAssetFaultReportedAssignedDowntimeImpactStatus
BT-26-034Mindray BS-480Reagent probe crash · error E-102213 Oct 05:40 · night MLTPremier Agencies (CMC)6.0 hChem TAT +2.4 h · backup analyserUnder repair
BT-26-031Autoclave 75 LDoor gasket leak · cycle failed10 Oct 14:10Zeeshan Raza71 hUsing tabletop autoclave · PR-DHA-26-0320Waiting parts
BT-26-029Pharmacy vaccine fridgeTemperature excursion during power cut10 Oct 20:00Zeeshan Raza0.7 hStock quarantined then releasedResolved
BT-26-027USG machineProbe cable damage02 Oct 12:00Imaging vendor7.8 h11 scans rescheduledResolved
Asset classMethodRateCostAccum. dep.Book valueQ3 charge
Lab analysersStraight line20%18,600,0007,440,00011,160,000580,000
Imaging equipmentStraight line15%14,200,0004,260,0009,940,000380,000
Clinic & day-care equipmentReducing balance15%2,100,000800,0001,300,000160,000
Furniture & ITReducing balance10–30%13,900,0005,330,0008,570,000—
Total48,800,00017,830,00030,970,0001,120,000

The quarterly charge posts as JV-26-10-0011 (awaiting approval in Accounts). Tax depreciation rates are kept separately [Verify].